Mako Total Hip

instrument prep: tray layouts

Mako Total Hip

instrument prep: VIZADISC assembly

VIZADISC assembly is performed by the sterile staff. These images show the arrays for Mako Total Knee. VIZADISC assembly is the same for all procedures.

Remove VIZADISC from the box and remove sterile top sheet.

The VIZADISC is arranged in patterns that exactly match the geometry of the arrays and the probes.

Place the chosen arrays over the appropriate layout.

Mako Total Hip

instrument prep: VIZADISC assembly

VIZADISC assembly is performed by the sterile staff. These images show the arrays for Mako Total Knee. VIZADISC assembly is the same for all procedures.

With the VIZADISC tray on a stable platform, press the array onto the VIZADISC.

Press each VIZADISC until a distinct snap is felt.

Lift the array from the tray at an angle, ensuring the VIZADISC is securely attached.

Visually inspect to ensure there are no gaps between the array and the VIZADISC.

no gap

Mako Total Hip

instrument prep: VIZADISC assembly

This demo shows the arrays for Mako Total Knee. VIZADISC assembly is the same for all procedures.